Labmate LMLPC-A200 Liquid Particle Counter

Labmate-LMLPC-A200-Liquid-Particle-Counter-PRODUCT

Giriş

Liquid Particle Counter LMLPC-A200 is configured with a light resistance method for accurate particle counting within the particle size range of 0.8µm to 600 µm in liquids. Includes a high-precision laser sensor that maintains stability at sampling speeds of 5 to 80mL/min. Ensures stable performance with low noise and high resolution. Equipped with viscosity, moisture, and temperature sensor modules for detailed analysis.

Xüsusiyyətlər

  • Offering a wide test range to accommodate various fluids and viscosities
  • Features an air purification system to prevent test contamination
  • Designed for testing hydraulic oil, lubricating oil, anti-fuel oil, insulating oil, and turbine oil
  • Adjustable cups accommodate testing requirements across different industries
  • Improved dual interface design ensures easy and convenient operation
  • Appropriate for inspecting products with different viscosities
  • Featuring multiple calibration curves compatible with international standards
  • Integrated data analysis system for comprehensive test data from a single test

Spesifikasiyalar

Model nömrəsi LMLPC-A200
İşıq mənbəyi Yarımkeçirici lazer
Hissəcik Ölçüsü Aralığı 0.8 µm-dən 600 µm-ə qədər
Detection Channels 8 to 64 channels are optional
Həssaslıq 0.8µm or 3µm
Qətnamə  
Sample Detection Viscosity ≤650cSt
Sampling Volume 0.2 to 1000mL, interval 0.1mL
Sampling Dəqiqliyi ±0.5%-dən daha yaxşı
Sampling Sürəti 5 to 80mL/min
Enerji təchizatı 110 to 245V AC,50Hz,70W
Dimensions (L × D × H) 340 × 410 × 650 mm
Xalis Çəki 25 Kq

Tətbiqlər

Liquid Particle Counter LMLPC-A200 plays a vital role in pharmaceuticals and semiconductor manufacturing, enabling the counting of suspended particles in liquids for product quality and compliance

Alətin Girişi

Sistem tərkibi

The instrument is made up of the following components:

  1. The main instrument is the primary device or mechanism designed to perform the core function.
  2. The pump is a secondary component that assists in the instrument’s operation by moving fluids or gases as needed.

Ön Panel 

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-1

  1. LCD displey: Indicates the menu touch interface realizes the operation of the instrument.
  2. High-pressure precision syringe: Used by sampuzanmaq.
  3. Təzyiq ölçən: Indicate the pressure value in the dome.
  4. Pressure valve: Adjust the maximum pressure value.
  5. Vacuum gauge: Indicate the vacuum value in the dome.
  6. Vacuum valve: Adjust the maximum vacuum value.
  7. Sampling needle: s üçün istifadə olunurampuzanmaq.
  8. Qübbə: Sealed with the pressure or vacuum.
  9. Dəstək: Operates the opening and closing of the dome.
  10. Inspection desk: To place the sample during the measurement.
  11. Printer: Print the measurement data.

Back Panel 

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-2

  1. Emergency pressure relief button: Release the Pressure in the dome when emergency. (Note: Besides the instrument)
  2. Müşahidə pəncərəsi: Observe the filter combinations.
  3. Serial port: Be connected with the computer to transmit data.
  4. USB port: USB disk storage port.
  5. Sıfırlama düyməsi: Reset the instrument. (Reserved function)
  6. Giriş həcmitage & switch: Connect with the power.
  7. Pump input voltage: Supply power for the pump.
  8. Outlet of liquid drain tubing: Connect with the drain tubing.
  9. Outlet of vacuum tubing: Connect with the vacuum outlet of the pump.
  10. Outlet for pressure tubing: Connect with the pressure outlet of the pump.
  11. Outlet sewage: Connected with sewage discharge device.

Quraşdırma

Quraşdırma mühiti 

  1. The instrument should be placed in an environment with relatively stable air flow less dust, and far away from electromagnetic interference sources.
  2. Suitable working temperature: 10~35℃, relative humidity: ≤80%.
  3. The power supply of the instrument should comply with AC100~240V.
  4. The grounding end of the instrument power socket should be equipped with a reliable grounding wire to ensure personal safety and reliable instrument operation.

Connect with Pump 

  1. Connect the outlet of pressure which is the back of the instrument with the pump use tube.
  2. Connect the outlet of the vacuum which is the back of the instrument with the pump use tube.

Install the High-Precision Syringe 

  1. Before the operation, the lifting arm should be in a lower position. If it is not, turn on the instrument then run the lifting arm down.
  2. Take the syringe from the packing case. First, make connecting hole under the syringe set on the lifting arm and screw the retaining screw. (Note: Don’t lock the screw, see figure).Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-3
  3. Pull up the syringe vertically, aligning the upper end of the syringe with the three-way valve, and screw the syringe carefully (see figure).Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-4
  4. Lock the screw (see figure).Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-4

İş prinsipi

Theory of the Instrument (Sensor)

The theory is described as follows:
The sensor is composed of a glass slit and a laser beam. The Laser beam is profiled for optimal illumination and responsiveness. The detector gathers the laser beam to produce a current. When the sample with particles through the glass slit from the bottom up due to negative pressure, the particle block part of the laser beam. This block leads to the detector’s output current being changed, the value of the changed current is proportional to the particle’s projected area. The formula is as follows:

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-6

  • Ip- Blocked current(µA)
  • Io – Normal current(µA)
  • A – Area of the glass slit(µm2)
  • a – Projected area of particle(µm2)

Working Diagram of the Instrument
The working diagram of the instrument is as follows:

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-7

  1. Alət s olduqdaampling, the three-way valve is turned to the sampling position, then the high-pressure precision syringe moves, the sample passes through the sampling needle through the sampling slit, and the particles in the sample block the laser light source, causing the photoelectric receiving end to produce electrical signals, the electrical signals through the amplification collection system to determine the size of the particles and count, and then transmitted to the single chip microcomputer.
  2. MCU control system, display, printing, and keyboard control. After sampling is completed, the three-way valve turns to the drain position, the high-pressure precision syringe moves, and the sample is discharged into the waste liquid (recoil) cup.
  3. During the recoil, the three-way valve turns to the drain position, the high-pressure precision syringe moves, and the recoil liquid is flushed from the waste liquid (recoil) cup to the injection slit to achieve the effect of cleaning the injection slit.

Əməliyyatlar

Instrument A Usage

Menyu Quruluşu 

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-8

Menu Button Description
In the menu, there will be a keyboard and some fixed functions of the keys, now it is explained.

Açar lövhə

  • OK: Change the data and exit the keyboard;
  • ESC: Do not change the data and exit the keyboard.

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-9

Əsas Təlimatlar

  1. Sidebar keys: Users can switch between management, measurement, setup, calibration, data, maintenance menus, and the stirring speed adjustment button. There are secondary menus for management, setup, calibration, and data. Click 》 to exit the secondary menu.
  2. ESC: Not save the settings and return to the previous menu;
  3. OK: Parametrləri yadda saxlayın.

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-10

İdarəetmə 

Daxil ol

After the instrument is powered on, the login interface is displayed. Enter the user’s name and the password through the keyboard. If the username and password are correct, the main interface will be entered. The name of the administrator is admin, the initial password is 0.

Qeyd: Nothing can be done without logging in!

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-11

Parolun dəyişdirilməsi

Click the “MGT. and PWD MOD.” button, then enter the interface of modify password. Enter the initial password, new password and confirm the password, and click the save button to save the new password.

Operator İdarəetmə

Qeyd: This function can only be used after logging in as an administrator.

  1. Click the “MGT. and User MGT.” button, then enter the interface of user management.
  2. After logging in to the system, the administrator can set some operators and permissions on the System Settings page. The user can set the following permissions: Measure set, System set, Calibrate and delete data. (Maximum 5 users)Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-12
  3. The right side of the display area displays the names of all current users, and the left side displays the user names and user permissions.
  4. Set up new user: Enter the corresponding information in the “User Name”, “Password” and “Confirm password” input boxes in the left display area. Choose the permission content in the permission checkbox and click Save.(Note: Maximum 8 characters for username and maximum 8 characters for password)
  5. Delete an existing user: Click the right display area to select the user’s name to be deleted, and click the DEL. button and click the Delete button again.
  6. Modify existing user permissions: Click the user’s name on the right, take “operator” as an example, and the user name “operator” and the permissions of the user “operator” are displayed on the left. Choose the permissions checkbox to change the user’s permissions; Click the “SAVE” button to save the changes to the user’s “operator” permission.

Quraşdırma
In the main menu press the SET button to display the Setup main menu. Meas. SET, Size SET, Out. SET and SYS. SET.

Ölçmə Quraşdırma

In the Setup main menu, click MEAS. button to display the measurement setup menu.

  1. Predicted volume: Volume used for pipeline flushing before test, not counted. For automatic testing, only one predictive flush is performed before a set number of tests. For manual testing, a predictive flush was performed before each test.
  2. Samphəcmi: 0.2~6000mL test will be carried out according to the set test volume.
  3. Repetition: 1~9 times, the test will be carried out according to the set test times.
  4. Positive pressure: 0~800kPa, the automatic test will be by the set pressure value.
  5. Vacuum pressure: 0~80kPa, degassing will be done according to the set pressure value during the automatic test.
  6. Duration: 0~59s, the automatic test will maintain the vacuum pressure according to the set time.
  7. Time interval: The interval time of each test in automatic testing; Sample speed: the flow rate during sampuzanmaq.
  8. Drain speed: The flow rate during drainage. Click the corresponding selection box to set whether the first data is included in the mean value, automatic storage, automatic printing, automatic data uploading, and data storage mode (local storage, USB disk storage).

Ölçü Quraşdırma

In the Setup main menu, click the SIZE SET button to display the Size setup menu. In this menu, the sizes of custom measurements can be set. The input particle size should be within the set range of particle size, otherwise, when saving, a prompt “Setting out of range! Please reset”. (Maximum 50 custom channel sizes can be entered)

Çıxış Quraşdırma

In the Setup main menu, Click OUT. SET button to display the output setup menu. In this menu, the output contents and the sample messages can be set, such as Sample Name(20 words Max), and Batch Number(10 letters or numbers Max). Click the corresponding text box to set up another item of output (Include “Sample Name”, “Batch Number”, “Sample Volume”, “Repetition”, “Test Time”, “Tester”, “Data”, “Mean” and “Level”). Finally, click the SAVE button to save all settings.

Sistem Quraşdırma

In the Setup main menu, press SYS. SET button to display the System setup menu. The user can modify the instrument system language and time.

Ölçmə
In the main menu press MEAS. button to display the Measurement Menu. All the measurement parameters can be set in the Measurement setup menu (Measurement Setup). There are two modes of measurement: Auto and Manual. The user can switch the test standard through the drop-down arrow, and set whether to select the automatic test mode. After the test, users can switch different standards to view, print, and store test data. When automatic testing is not selected, manual pneumatic operations such as pressure, pause, vacuum, and exhaust can be carried out before, during, and after testing. When the automatic test is selected, the instrument automatically carries out pressure pause, vacuum and other operations according to the pressure set value in the test setting.

The channel particle sizes of the 13 standard tests are as follows:

  • NAS1638: 5-15µm, 15-25µm, 25-50µm, 50-100µm, >100µm
  • SAE4059F: 5-15μm, 15-25μm, 25-50μm, 50-100μm, >100μm
  • SAE4059CPC: >4μm(c), >6μm(c), >14μm(c), >21μm(c), >38μm(c), >70μm(c) ISO4406: >4µm(c), >6µm(c), >14µm(c)
  • SAE749D: 5-10μm, 10-25μm, 25-50μm, 50-100μm, >100μm
  • Custom test: No. of channels and channel particle size in custom settings.

Data
In the main menu press the “DATA” button to display the saved data Menu. The data that is saved in measurement can be reviewed and queried in this menu.

Data Review

In the DATA menu, Click the “REVIEW” button to display the Data Review menu. In this menu, the measurement data can be reviewed. Users can switch standards and view data and means for different test times under each group of data. Also, can delete the current group, clear all data, export the current group data, print the current group data, upload the current group data, and view the current group data information.

Data Sorğu

In the DATA menu, click the “QUERY” button to display the data query menu. In this menu, click the textbox to edit the measurement time to query. Click the “QUERY” button to filter the saved data, the filtered data will be displayed. Turn the page by pressing the PgUp and PgDn buttons, and click the number of groups to display the specific test data and level on the right side of the interface. Click the left and right arrows to switch the number of times in the current group. Click the PRINT button to print the data of the current group. Click the EXP. button to export the group data to a USB drive. Click the “UPLOAD” button to upload the group data to the PC side. Click the “INFO” button to view the details of the current group data.

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-19

Baxım
In the main menu click the “MAINTENANCE” button to display the maintenance menu. In this menu, the Flush and Back Flush can be operated. This is very important to the maintenance of the instrument.

Users can read the prompt information on the maintenance interface carefully and operate according to the prompt information. Users can choose whether to carry out a cleaning test. If you choose the cleaning test, the instrument will stop cleaning after judging that the actual particle concentration meets the requirements according to the calibration curve, particle size, and corresponding particle concentration. If the cleaning test is not selected, stop cleaning when the cleaning volume is reached.

Qeyd: Ensure to make it empty before backflushing!

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-20

Kalibrləmə
In the main menu click the “CAL” button to display the Calibration Menu. Noise test, Size calibration, Volume calibration, and channel Resolution can be carried out. Click the corresponding button to enter the corresponding interface.

Səs-küy Testi

In the Calibration main menu, click the “NOISE TEST” button to display the Noise Level menu. The noise test is used to check the electrical noise of the instrument. The threshold level of the minimum calibration particle size can be set according to the noise test value.

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-21

Click the TEST button to start the Noise test. Fill the sensor with clean and non-flowing liquid, start the instrument to count the 60s, adjust the threshold level until the measured value meets the range, and then the threshold level is the noise level of the instrument.

Size Calibration

In the Calibration main menu, click SIZE CAL. button to display the SIZE calibration menu. The user can enter the setting interface through the SET button, then can set the calibration volume, speed, and calibration curve and can calibrate up to three curves of ACFTD, ISOMTD, and custom.
Take the manual calibration of the ISOMTD curve as an example: In the particle size column (μm), the particle size value required for calibration is entered, and the corresponding threshold level (mV) is entered after this particle size value. Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-22

Click the “CAL” button, the instrument will automatically extract the calibration solution, and give the measured value of each particle size per unit volume after the end. According to the deviation between the measured value and the calibration liquid standard value, the user can increase (or decrease) the particle size threshold level and test again until the deviation between the measured value and the calibration liquid standard value meets the corresponding national standard calibration requirements. Click the “SAVE” button to complete the instrument calibration operation.

Həcmi Kalibrləmə

In the Calibration main menu, click the “VOLUME CAL” button to display the Volume calibration menu. The sampling volume can be calibrated, and the calibrated volume is fixed at 5mL.

Kalibrləmə addımları aşağıdakı kimidir:

  1. Take a clean and dry beaker and fill it with about 100ml purified water (or other calibrated volume solution), taking care that there are no water droplets on the outer wall of the beaker. They were weighed on a balance and the values were recorded.
  2. Clean the inspection table to make it clean and dry.
  3. Place the beaker containing pure water on the inspection table and move the inspection table upward so that the needle is submerged in the pure water.
  4. Click the “CAL” button on the volume calibration interface, and the instrument will take sampavtomatik olaraq.
  5. s sonraampling, the instrument enters the volume difference input interface. The beaker was removed, weighed again on the balance, and the data recorded.
  6. The difference between the two weighing results is converted into volume and then input to the volume difference input interface of the instrument.
  7. Click the SAVE button to complete the volume calibration. If the injection volume is not accurate, the sampling operation can be repeated several times. If the volume calibration fails, you can initialize the volume parameters by pressing the RESET button.

Kanal Çözünürlüğü

In the Calibration main menu, click CH. RESOLUTION button to display the channel resolution menu. Enter the threshold value, click the TEST button, and wait for the instrument to complete the test of channel resolution.

Qeyd: This item can only be operated by the manufacturer or unit of measurement.

Stir Setup

The mixing speed can be adjusted directly through the + and – buttons on the interface, and the adjustment range is 12 gears.

Test əməliyyatı

  1. Connect to an electrical outlet and turn on the instrument. The instrument will conduct a self-test. After the self-check is completed, the login page is displayed. Enter the correct user’s name and password, and then should be preheated for 10 minutes before measuring.
  2. Kimi yerləşdirinampling bottle containing petroleum ether solution (grade 6 or below, or other suitable cleaning solution) on a sampling table and close the air pressure chamber (Dome operate), enter the flushing interface, set the flushing times, and carry out flushing.
  3. Dome operation:
    The operation of sealing the dome
    • A hand holds the handle and rotates the handle to align the two signs.
    • The other hand moves up the test sets to the stop position.
    • Rotate the handle to the stop position to seal the dome.
      The operation of open the dome 
    • A hand holds the handle and the other hand holds the test sets.
    • Rotate the handle counterclockwise until the two signs align.
    • Move the test sets down slowly to open the dome.
      Xəbərdarlıq: Before opening the dome, make sure there is no pressure in the dome. If there is pressure in the dome, press the manual pressure relief valve to relieve the pressure manually!
  4. Setup the instrument, set the Test times, Sample volume and others (Measurement Setup), print format (Output Setup), channel particle size and channel number (Custom Setup), system time (System Setup), etc.
  5. Disperse the samples to be tested: tighten the cap and put it into the ultrasonic cleaning tank (with power of at least 4000W/m2 ) to vibrate for at least 1 minute. The liquid in the ultrasonic cleaning tank should be the same as the liquid level of the sample bottle to be tested.
  6. After the ultrasonic vibration, shake the sample bottle vigorously by hand for 5 minutes. Then again ultrasonic wave mu vibration or the use of negative pressure degassing, time should be as short as possible. Let stand for 2-3 seconds.
  7. Enter the measure interface to select the standard test type displayed, and click the TEST button to conduct the test. If the test sample is viscous and cannot be tested properly, pressure can be applied to assist the test during the test (Measurement).
  8. At the end of the test, the previous test data of the current group can be viewed, it can be stored, and print test data.
  9. After completion of the test, perform at least 10 times cleaning operations with petroleum ether flushing solution or other suitable solution (Maintenance). After flushing, test the next sample or shut it down.

Baxım

  1. The sensor assembly shall not be disassembled by itself under any circumstances.
  2. Sampling window and case should be kept clean and dry to prevent the inspection liquid from corroding the instrument components and case.
  3. When the instrument is not in use, the power supply of the instrument should be turned off to extend the usage life of the laser parts.
  4. The high-pressure syringe should be cleaned regularly to prevent the high-pressure syringe from being contaminated and causing inaccurate measurement volume. Remove the high-pressure syringe before cleaning, and then use methanol, dichloromethane, acetonitrile, acetone, and other solutions for cleaning.
  5. The instrument often detects the inspection liquid with large viscosity or large particle size, which is easy to cause the blockage of the slit of the injection glass. The performance is as follows prolonged injection time.
  6. In the process of sample injection, the speed of sample injection is slower than that of the high-pressure syringe, and a large amount of gas is pumped into the high-pressure syringe. The data is too large and unstable, etc., and the processing methods are as follows:
    Reverse flushing method: Backflushing to reverse cleaning in maintenance.
    Alternate cleaning method: Enter the flush interface to input the cleaning times. At the beginning of cleaning, remove the sample liquid level from the injection needle, and make the air and sample enter the injection glass slit alternately. Repeat several times to remove the substances blocking the injection glass slit and make it unblocked.
    Flushing with cleaning agent: Use a neutral cleaning agent or other suitable solvent for cleaning. Enter the flush interface and input the cleaning times to clean the slit and pipeline of the sample injection glass, which is used to clean the remaining samples, oil stains, and impurities in the pipeline. Rinse the pipeline again.

Əlavə

Sensorun çıxarılması

When the instrument gives out false results, the sensor blocking can be taken into consideration. When the block is exited, flush the instrument with flush liquids times. If the problem cannot be solved, the operation of removing the sensor can be done. Then Use suitable tools to clean the blocked sensor. The procedure for sensor removal is as follows:

  1. Flush the instrument times and drain out.
  2. Open the lid of the instrument: Use a screwdriver to screw the two retaining screws, then lift the lid.
  3. Disconnect the data line: Use a screwdriver to screw the retaining screw to off the data line.
  4. Remove the top joint of the sensor: A hand holds on the sensor; the other hand uses a spanner to unlock the joint anticlockwise.
  5. Remove the sensor: A hand holds the sensor, and the other hand uses the spanner to unlock the lower-end joint.

Air Clean Combination Treatment

  1. Air filter combined drain out
    • When the instrument is in a certain period, especially in the air humidity of the environment, the air filter combination may be the remnants of a certain amount of water.
    • Through the observation window back of the instrument to see whether it needs to drain out.
    • When the drain-out operation is needed, firstly, exhaust the pressure or vacuum in the dome, then power off the instrument.
    • Secondly, Use the screwdriver to unlock the retaining screw of the view window, and put a vessel in the lower end of the air filter.
    • Thirdly, screw the black lid to drain.
      Qeyd: Make sure there is no water leakage into the inside of the instrument.After the drain-out operation, screw the black lid tightly to avoid leakage.)
  2. Replace the dryer
    • Firstly, remove the back panel of the instrument.
    • Secondly, screw the retaining screw of the drying cylinder, pull off the pressure tube on the drying cylinder, then screw off the sealing lid on the other side.
    • Replace the dryer, then screw the sealing lid and the pressure tube. Screw the drying cylinder tightly using the screwdriver.
    • Finally, fix up the back panel of the instrument.
  3. Replace the filtration module
    When the pressure pump is inefficient, replacing the filtration module should be considered. The procedure is as follows:
    • Firstly, remove the back panel of the instrument using the screwdriver.
    • Secondly, pull off the pressure tube of the filtration module, then screw the retaining screw of the filter, and replace the filter.
    • Thirdly, install the filtration module with the reverse procedure.
      Qeyd: When replacing the filter, make sure both O-rings are placed in the middle.

Təmizlik sinfi

Təmizlik Sinif Size/µm
5-15 15-25 25-50 50-100 >100
00 125 22 4 1 0
0 250 44 8 2 0
1 500 89 16 3 1
2 1000 178 32 6 1
3 2000 350 63 11 2
4 4000 712 126 22 4
5 8000 1425 253 45 8
6 16000 2850 506 90 16
7 32000 5700 1012 180 32
8 64000 11400 2025 360 64
9 128000 22800 4050 720 128
10 256000 45600 8100 1440 256
11 512000 91200 16200 2880 512
12 1024000 182400 32400 5760 1024
Sinif Less than or equal to (Particles/100ml) > %
>0.5-1 >1-2 >2-5 >5-10 >10-25 >25-50 >50-100 >100-200    
00 800 400 32 8 4 1   AO  

 

AO

0 1600 800 63 16 8 2  
1   1600 125 32 16 3
2   250 63 32 4 1
3   125 63 6 2
4 250 125 12 3
5 500 250 25 4 1
6 1000 500 50 6 2 1 0.000032
7 2000 1000 100 12 4 2 0.000064
8 4000 2000 200 25 6 3 0.000125
9 8000 4000 400 50 12 4 0.00025
10 16000 8000 800 100 25 5 0.0005
11 31500 16000 1600 200 50 10 0.001
12 63000 31500 3150 400 100 20 0.002
13   63000 6300 800 200 40 0.004
14         125000 12500 1600 400 80 0.008
15       25000 31500 800 160 0.016
16       50000 63000 1600 315 0.032
17         125000 3150 630 0.064
Təmizlik Sinif 5-15 um 15-25 um 25-50 um 50-100 um >100um
6-14 um(c) 14-21

um(c)

21-38

um(c)

38-70

um(c)

>70 um(c)
00 125 22 4 1 0
0 250 44 8 2 0
1 500 89 16 3 1
2 1000 178 32 6 1
3 2000 356 63 11 2
4 4000 712 126 22 4
5 8000 1425 253 45 8
6 16000 2850 506 90 16
7 32000 5700 1012 180 32
8 64000 11400 2025 360 64
9 128000 22800 4050 720 128
10 256000 45600 8100 1440 256
11 512000 91200 16200 2880 512
12 1024000 182400 32400 5760 1024
Təmizlik Sinif >1µm >5µm >15µm >25µm >50µm
>4µm(c) >6µm(c) >14µm(c) >21µm(c) >38µm(c)
000 195 76 14 3 1
00 390 152 27 5 1
0 780 304 54 10 2
1 1560 609 109 20 4
2 3120 1217 217 39 7
3 6250 2432 432 76 13
4 12500 4864 864 152 26
5 25000 9731 1731 306 53
6 50000 19462 3462 612 106
7 100000 38924 6924 1224 212
8 200000 77849 13849 2449 424
9 400000 155698 27698 4898 848
10 800000 311396 55396 9796 1696
11 1600000 622792 110792 19592 3392
12 3200000 1245584 221584 39184 6784
/ml Kod
-dən böyükdür Daha az və ya bərabərdir
2500000   >28
1300000 2500000 28
640000 1300000 27
320000 640000 26
160000 320000 25
80000 160000 24
40000 80000 23
20000 40000 22
10000 20000 21
5000 10000 20
2500 5000 19
1300 2500 18
640 1300 17
320 640 16
160 320 15
80 160 14
40 80 13
20 40 12
10 20 11
5 10 10
2.5 5 9
1.3 2.5 8
0.64 1.3 7
0.32 0.64 6
0.16 0.32 5
0.08 0.16 4
0.04 0.08 3
0.02 0.04 2
0.01 0.02 1
0.00 0.01 0
Təmizlik Sinif Particle size range (μm)
5~15 15~25 25~50 50~100 >100
0 2700 670 93 16 1
1 4600 1340 210 28 3
2 9700 2680 380 56 5
3 24000 5360 780 110 11
4 32000 10700 1510 225 21
5 87000 21400 3130 430 41
6 128000 42000 6500 1000 92

Changing Printer’s Thermo-paper Roll

Prosedur aşağıdakı kimidir: 

  1. Press the open-door button to open the printer’s door. Take out the rest paper roll.Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-23
  2. Changing the new paper roll followed the right picture.
    Ehtiyatlı olun! Take note of the right direction, otherwise, the paper will not be printed.Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-24
  3. Close the printer’s door as the picture shows.
    Qeyd: before closing the door, make sure, the paper is out of the door a little, then the printer will print correctly.

Labmate-LMLPC-A200-Liquid-Particle-Counter-FIG-25

ƏLAQƏ

Labmate Scientific Inc
E-poçt: info@labmate.com | Websayt: www.labmate.com

Sənədlər / Resurslar

Labmate LMLPC-A200 Liquid Particle Counter [pdf] Təlimat kitabçası
LMLPC-A200 Liquid Particle Counter, LMLPC-A200, Liquid Particle Counter, Particle Counter
Labmate LMLPC-A200 Liquid Particle Counter [pdf] Təlimat kitabçası
LMLPC-A200, LMLPC-A200 Liquid Particle Counter, LMLPC-A200, Liquid Particle Counter, Particle Counter, Counter

İstinadlar

Şərh buraxın

E-poçt ünvanınız dərc olunmayacaq. Tələb olunan sahələr qeyd olunub *